MAN 01/EN/03 CRONO PAR 50 06/13
Auto
BUTTONS SETTINGS DISPLAY
PUMP SET TO OFF
• Pump set to OFF and L0
• Visualisation of the time
• Enabling the setting of the time
• Set the minutes
• Set bolus dose volume (from 0.00 to 2.00 ml)
• Setting the interval between the bolus doses
(from “no,Lt” to 24 hours)
• Setting the ow rate F1 (from 0.05 to 5.00 ml/h)
• Set the F2 ow rate (from 0.05 to 5.00 ml/h)
• Set the F3 ow rate (from 0.05 to 5.00 ml/h)
Assigning the ows over the daily 24-hour period
• Setting the partial volume (from 1 to 50 ml)
(see table SETTING THE PARTIAL VOLUME)
• Pump set to OFF (after 20 seconds)
• Decrease/increase stored value
/
SECTION 13
and
PROG
PROG
PROG
PROG
PROG
PROG
PROG
ml/h
PROG
ml/h
PROG
ml/h
(5 sec)